Intera 1.5T MRI System-Main Unit(2014-05-08)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
Intera 1.5T MRI System-Main UnitINTERA 1.5T

Introduction

Recalled Products

Intera 1.5T MRI System-Main Unit

Reason

Philips has become aware that the cross reference lines of transversal images from spinal scans are displayed in an incorrect position on a fused image which has been generated through mobiview post processing. The cross reference lines are correctly displayed on the unfused stations.  The problem can occur when the field of view of the clinical scan stations is below (more towards the feet) the light visor position. In such situation an error is made in calculating the position information. This causes incorrect positioning of cross-reference lines on fused sagittal images.