Single Width Airway Module(2014-05-21)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable
Manufacturer
GE Healthcare Finland OY
Kuortaneenkatu 2,
Helsinki
00510
FINLAND
Single Width Airway ModuleE-MINIC-00

Introduction

Recalled Products

Single Width Airway Module

Reason

GE has found that there can be a failure of the CO2 detector in the single-width airway and extension modules. Failure of the CO2 detector in single-width airway and extension modules may cause a slow continuous decrease of measured Et-/FiCO2 values. Incorrect et CO2/FiCO2 value may impair clinical decision making for both mechanically and spontaneously ventilated patients because of incorrect low CO2 values.