Architect I System - EBV VCA IgM controls (2014-05-14)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
38938LI00Architect I System - EBV VCA IgM controls 
Manufacturer
Abbott GmbH and Co. K.G.
Max Planck Ring 2
Wiesbaden
65205
GERMANY
3P66-10

Introduction

Recalled Products

Architect I System - EBV VCA IgM controls 

Reason

During stability testing a significant signal drop for the positive control of architect EBV VCA IgM control kit compared to the release results was recognized, indicating a loss in signal over time. Subsequent out of specification low results for the positive control of Architect EBV VCA IgM control kit was documented confirming above observation of a loss in signal over time.