Health Products

Vitros Chemisty Products HSCRP Reagent (2014-05-12)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
1508-28-3487
Manufacturer
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
Vitros Chemisty Products HSCRP  Reagent (6801739

Introduction

Recalled Products

Vitros Chemisty Products HSCRP  Reagent 

Reason

Calibrator values were released using a reference method that was undergoing validation for a new lot of European Reference Material (ERM). This validation process was not complete prior to calibrator value assignment for Vitros chemistry products HSCRP reagent (Product Code 6801739), Generation (Gen) 28.