Health Products

X Series/PropaqMD/PropaqM monitor/defibrillator running software version 02.10.02.00 or higher (2014-03-10)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Zoll Medical Corporation
269 Mill Road
Chelmsford
01824
Massachusetts
UNITED STATES
X Series/PropaqMD/PropaqM monitor/defibrillator running software version 02.10.02.00 or higher610-0120011-01

Introduction

X Series/PropaqMD/PropaqM monitor/defibrillator running software version 02.10.02.00 or higher 

Reason

The device displayed an "internal error reset required" message and executed a system reset. This could occur while monitoring patients or may enter a continuous "System Reset" loop.