Zeus IE Anesthesia Workstation (2014-05-30)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Draeger Medical GMBH
Zeus IE Anesthesia WorkstationMK04000

Introduction

Recalled Products

Zeus IE Anesthesia Workstation

Reason

The device is designed to generate an audible and visual alarm when the measured value for spo2 falls below the alarm threshold set for the spo2 low alarm by the clinician. If the measured value continues to fall, the device is designed to continue to generate a spo2 low alarm. But, for devices with software version 1.03, if the sp02 measured value drops below the spo2 low alarm threshold by more than 10 percent, the audible and visual alarm for sp02 low will stop. If the value measured rises again, the alarm will be generated again as soon as the measured value is ten percent or less below the spo2-low alarm threshold. All displayed measurements of values and spo2 curves are performing as designed at any time.