Kodak Directview DR 9000 System (2014-06-02)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
102
Manufacturer
Carestream Health Inc.
150 Verona Street
Rochester
14608
New York
UNITED STATES
Kodak Directview DR 9000 System851 9530

Introduction

Recalled Products

Kodak Directview DR 9000 System

Reason

Carestream has identified that in the case of a control system failure, the DR 9000 u-arm may move downward unexpectedly when the radiology technologist is positioning the equipment using the "up" or "down" device control buttons. This downward movement occurs at three times the normal rate of speed and will continue until the device control button is released, the emergency stop button is activated or the equipment bump sensor contacts an object. There is a potential for injury if this issue occurs for patients positioned under the u-arm for decubitus exams.