Triage B-type Natriuretic Peptide (BNP) test for the Beckman Coulter Immunoassay Systems
98202
Introduction
Recalled Products
Triage B-type Natriuretic Peptide (BNP) test for the Beckman Coulter Immunoassay Systems
Reason
This issue is limited to a single lot of the product. In a portion of lot K328163, the bottle labeled as calibrator level 5 actually contains calibrator level 4. This results in a calibration failure when attempting to calibrate using a calibrator kit which contains the affected level 5 calibrator. Although not all kits from this lot are impacted, we cannot determine which specific calibrator level 5 bottles in this lot are affected. Therefore it is recommended that customers stop using all products from this lot and discard any unused product.