Walkaway Plus System - 40 - Instrument (2014-06-05) Jun 5, 2014 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Companies Brand name Model or catalog number All serial numbers distributed Manufacturer Siemens Healthcare Diagnostics Inc. 500 GBC Drive, Mailstop 514, PO Box 6101 Newark 19714-6101 Delaware UNITED STATES Walkaway Plus System - 40 - Instrument B1018-283
Recalled Products
Walkaway Plus System - 40 - Instrument
There is a potential failure with access door hinge. The failure can cause the springs to dislodge and be ejected towards the operator causing a potential safety issue.