Somatom Emotion 16-slice Configuration (2014-06-17)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
S/N 85828
Manufacturer
Siemens Shanghai Medical Equipment Ltd.
278 Zhouzhu Road,
A. Somatom Emotion 16-slice Configuration10165977

Introduction

Recalled Products
  1. Somatom Emotion 16-slice Configuration 

Reason

Siemens has recently detected a rubber component with unfavorable ageing properties on the Somatom Emotion 16-slice Configuration. The mechanical function of this rubber damper, which is situated within the mounting assembly of the rotation motor, may potentially degrade over time and may impact stability.