Introduction
Recalled Products
A) Proteus XR/A General Purpose Radiographic System
B) Revolution XR/D Digital Radiographic System
C) Proteus XR/A General Purpose Radiographic System
D) Revolution XR/D Digital Radiographic System
Reason
If a device is improperly re-installed, the collimator may not be properly seated and may loosen and fall. Such an event may occur immediately following re-installation during a service maintenance activity or may also occur a few months after a maintenance and continued use. A fall of a collimator while the system is in use could result in an injury to a patient or operator. There was a reported incident of a collimator fall which caused a serious patient injury.