Proteus XR/A General Purpose Radiographic System and Revolution XR/D Digital Radiographic System (2014-06-12)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer.D) Revolution XR/D Digital Radiographic System
Manufacturer
GE Medical Systems LLC
3200 North Grandview Blvd.
Waukesha
53188
Wisconsin
UNITED STATES
GTC14064PC-2
More than 10 numbers, contact manufacturer.B) Revolution XR/D Digital Radiographic System
Manufacturer
GE Hualun Medical Systems Co. Ltd.
No.1 Yong Chang North Road,, Beijing Economic Technological, Development Zone,
Beijing
100176
CHINA
GTC14064PC-2
More than 10 numbers, contact manufacturer.C) Proteus XR/A General Purpose Radiographic System
Manufacturer
GE Medical Systems LLC
3200 North Grandview Blvd.
Waukesha
53188
Wisconsin
UNITED STATES
2259988
More than 10 numbers, contact manufacturer.A) Proteus XR/A General Purpose Radiographic System
Manufacturer
GE Hualun Medical Systems Co. Ltd.
No.1 Yong Chang North Road,, Beijing Economic Technological, Development Zone,
Beijing
100176
CHINA
2259988

Introduction

Recalled Products

A) Proteus XR/A General Purpose Radiographic System
B) Revolution XR/D Digital Radiographic System
C) Proteus XR/A General Purpose Radiographic System
D) Revolution XR/D Digital Radiographic System

Reason

If a device is improperly re-installed, the collimator may not be properly seated and may loosen and fall. Such an event may occur immediately following re-installation during a service maintenance activity or may also occur a few months after a maintenance and continued use. A fall of a collimator while the system is in use could result in an injury to a patient or operator. There was a reported incident of a collimator fall which caused a serious patient injury.