Health Products

WaterLase iPlus Dental Laser System (2014-06-03)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
72120052
Manufacturer
Biolase, Inc.
4 Cromwell
Irvine
92618
California
UNITED STATES
WaterLase iPlus Dental Laser System7200854

Introduction

Recalled Products

WaterLase iPlus Dental Laser System

Reason

During use of an iPlus unit with software version 1.7, if an error occurs when in any procedure screen, the air setting will change to 1% and off. After the error is cleared, the air setting will erroneously remain at 1% and off.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime