Gemini TF16 PET/CT System - Main Unit (2014-06-18)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
7099
Manufacturer
Philips Medical Systems (Cleveland), Inc.
595 Miner Road
Cleveland
44143
Ohio
UNITED STATES
Gemini TF16 PET/CT System - Main Unit1535 679 83931

Introduction

Recalled Products

Gemini TF16 PET/CT System - Main Unit

Reason

Philips Healthcare received a report from a customer that during a system start-up operations (powers up) when the power of the Gantry and CT host has been powered off), the E-stop (emergency stop that halts all scanner and table motions and X-ray generation) failed to close.