Artiste MV System - Digital Linear Accelerator (2014-06-12)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
5414
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
Artiste MV System - Digital Linear Accelerator8139789

Introduction

Recalled Product

Artiste MV System - Digital Linear Accelerator

Reason

Siemens became aware of an error in the pressure/temperature compensation for dose monitor 2 for electron treatments. Only Digital Linear Accelerators (Linac) running control console software version 12 or 13 are affected. This error may result in a higher deviation of the dose rate in electron mode than specified. The dose monitor 1 controls the overall dose monitoring system and remains unaffected. The actual dose rate and the accumulated monitor units and thus the absolute dose are correctly displayed at the linac console.