Health Products

Various Linear Accelerators Oncor Systems (2014-06-03)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
3799
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
D) Oncor Impression IMRT Plus5857912
5037
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
B) Oncor Expression Digital Linear Accelerator7360717
5414
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
C) Artiste MV - Linear Accelerator - Main Unit8139789
2559
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
A) Mevatron Linear Accelerators01940035

Introduction

Recalled Products

A) Mevatron Linear Accelerators
B) Oncor Expression Digital Linear Accelerator
C) Artiste MV - Linear Accelerator - Main Unit
D) Oncor Impression IMRT Plus

Reason

Siemens became aware of an incident with a Linear Accelerator Oncor where an auto sequence was set up with a treatment management system. During the auto sequence the gantry collided with the patient's head while moving. The patient received a wound on the head that resulted in skin injuries but no further damage. The user provided a detailed report in which he described that a mistake was made when selecting the beams for auto sequence. The user error caused the unintended motion and hence the injury.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime