A) Mevatron Linear Accelerators
B) Oncor Expression Digital Linear Accelerator
C) Artiste MV - Linear Accelerator - Main Unit
D) Oncor Impression IMRT Plus
Reason
Siemens became aware of an incident with a Linear Accelerator Oncor where an auto sequence was set up with a treatment management system. During the auto sequence the gantry collided with the patient's head while moving. The patient received a wound on the head that resulted in skin injuries but no further damage. The user provided a detailed report in which he described that a mistake was made when selecting the beams for auto sequence. The user error caused the unintended motion and hence the injury.