840 Ventilator System - Main Unit (2014-06-10)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable
Manufacturer
Covidien LLC
15 Hampshire Street
Mansfield
02048
Massachusetts
UNITED STATES
840 Ventilator System - Main Unit4-840120DICU-EN

Introduction

Recalled Products

840 Ventilator System - Main Unit

Reason

Customer reports have included the GUI display screens (both upper and lower) becoming blank during patient use. This event interferes with the clinician's ability to view and/or change ventilator setting or clinician-set patient alarm parameters. The ventilator continues to provide respiratory support during these events. All ventilator alarms remain active and will activate if alarm limits are exceeded.