Nexdrive Micropositioning Drive Models ML-1000 and ML-2000
Manufacturer
Medtronic Inc.
710 Medtronic Parkway NE
Minneapolis
55432
Minnesota
UNITED STATES
MI-1000
Introduction
Recalled Products
Nexdrive Micropositioning Drive Models ML-1000 and ML-2000
Reason
Medtronic is initiating a voluntary recall for the Nexdrive micro-positioning drive (Model ML-1000 and ML-2000) due to the potential for misalignment of the z-stage scale. Using one of these devices for a procedure could result in the microelectrode being inserted to an incorrect target depth. Medtronic received one (1) report of a misaligned z-stage scale discovered during a deep brain stimulation (DBS) surgical procedure that noted the microelectrode was 5 mm short of the target depth. This resulted in multiple microelectrode insertions and a prolonged duration for the surgical procedure.