Health Products

OMNI-Lab - Laboratory Information System (2014-06-16)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
Version 12.10.0.129A. OMNI-Lab - Laboratory Information System 
Manufacturer
Omnitech Labs Inc.
215, Boul. Du Seminaire Sud, Suite 1305
Saint-jean-sur-richelieu
J3B 8W1
Quebec
CANADA
Not applicable

Introduction

Recalled Products
  1. OMNI-Lab - Laboratory Information System 

Reason

The result of an examination for parasites may be transferred to a microbiology requisition (culture). In other words, if a user selects a culture result for a parasitology requisition at the "identification" tab, then clicks in the "requisition number" field of the basic tab of the main screen of the microbiology module, and then enters a microbiology requisition number (or scans a requisition label using a barcode reader), the result of the parasitology requisition on the screen is in the microbiology requisition called to the screen. The microbiology requisition result is replaced with a parasitology result.