Health Products

LIFEPAK 1000 Defibrillator (2014-06-05)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer.A. LIFEPAK 1000 Defibrillator
Manufacturer
Physio-Control, Inc.,
11811 Willows Road North East,
Redmond
98052
Washington
UNITED STATES
99425-XXXXXX

Introduction

Recalled products
  1. LIFEPAK 1000 Defibrillator

Reason

Physio-Control has become aware of incidents where customers have attempted to use their LIFEPAK 1000 and the device has shut down due to a very low battery. A defibrillator in this scenario has the potential to fail to deliver a shock; with the result that therapy is not delivered. The operating manual may contribute to customers not replacing the battery when required. The readiness display may indicate a low battery charge when it should indicate a very low battery charge. Batteries at both low and very low star of charge must be replace with a fully charged battery.