Health Products

Intouch Bed Caster Stem (2014-07-02)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer. 
Manufacturer
Stryker Medical
3800 E Centre Avenue
Portage
49002
Michigan
UNITED STATES
Intouch Bed Caster Stem2130

Introduction

Recalled Products

Intouch Bed Caster Stem 

Reason

Stryker received complaints which allege that a caster has broken on the top of the caster stem where the brake rod inserts into the caster. If more than one caster on a unit is damaged, brake functions may be affected and result in reduced brake holding force. If only one caster on the bed is damaged, brake functionality continues to meet specifications.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime