Health Products

Kodak Directview DR 7500 system (2014-07-11)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
75000150
Manufacturer
Carestream Health Inc.
150 Verona Street
Rochester
14608
New York
UNITED STATES
Kodak Directview DR 7500 system 879 1345

Introduction

Recalled Products

Kodak Directview DR 7500 system 

Reason

Carestream has identified an issue at a customer site related to the DR 7500 Beta Assembly (X-ray tube and collimator) detaching from the overhead tube crane (OTC) telescope. It has been determined that as a result of an improperly assembled part which holds the beta assembly to the OTC telescope, an inspection of the part is needed, and in addition, a repair of the assembly may be required.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime