Health Products

Integra LED Battery Charger, Single Bay (2014-06-11)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Integra Burlington MA Inc.
22 Terry Ave.
Burlington
01803
Massachusetts
UNITED STATES
Integra LED Battery Charger, Single Bay95023

Introduction

Recalled Products

Integra LED Battery Charger, Single Bay 

Reason

Integra Lifesciences has identified through an investigation of complaints that some lot numbers of the Integra LED single bay battery charger (Catalogue No. 90523) may prematurely fail and will not charge the Integra LED battery as intended. There have been no reports of an injury or other adverse patient consequences associated with the complaints. Customers may have received an affected 90523 directly, or had it provided with one of the following: LED headlight systems (Catalogue No. 90520US) or battery kits (Ccatalogue No. 90530US).