Blood Grouping Reagent MTS A/B/D Monoclonal and Reverse Grouping Card
Manufacturer
Micro Typing Systems Inc.
1295 29th Ave. S.W.
Pompano Beach
33069
Florida
UNITED STATES
MTS080515
Introduction
Recalled Products
Blood Grouping Reagent MTS A/B/D Monoclonal and Reverse Grouping Card
Reason
Ortho Clinical Diagnostics (OCD) observed an increased number of complaints related to false positive and/or reactions containing haze in the Anti-A and buffered gel microtubes used for reverse ABO typing in the MTS A/B/D Monoclonal and Reverse Grouping Card lots listed above. Our investigation determined that the false positive reactions were due to broken gel beads in the microtubes of the MTS cards.