Portex Endotracheal Tube Holders, Sizes 2.5mm and 3.0mm
H4051
Introduction
Recalled Products
Portex Endotracheal Tube Holders, Sizes 2.5mm and 3.0mm
Reason
A trend in complaints of difficulty in passing a suction catheter or locking the ET tube holder in place was observed. Smith Medical's investigation determined that the fastener component and cylinder component on the ET tube holder could cause compression of the tracheal tube when assembled together. The investigation determined that in some cases, the fastener and cylinder could compress the tube enough to affect the tracheal tube inner diameter. Customer feedback suggested that difficulty passing suction catheter usually occurred when a size 6FR suction catheter (or larger) was used.