D) Vitros 3600 Immunodiagnostic System (Licence #80702)
680 2783
More than 10 numbers, contact manufacturer.
Manufacturer
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
E) Vitros 4600 Chemistry System
680 2445
More than 10 numbers, contact manufacturer.
Manufacturer
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
B) Vitros 5600 Integrated System (Licence #78571)
680 2413
More than 10 numbers, contact manufacturer.
Manufacturer
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
C) Vitros 5600 Integrated System (Licence #80701)
680 2413
36000130
Manufacturer
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
A) Vitros 3600 Immunodiagnostic System (Licence #78572)
680 2783
Introduction
Recalled Products
A) Vitros 3600 Immunodiagnostic System (Licence #78572)
B) Vitros 5600 Integrated System (Licence #78571)
C) Vitros 5600 Integrated System (Licence #80701)
D) Vitros 3600 Immunodiagnostic System (Licence #80702)
E) Vitros 4600 Chemistry System
Reason
OCD has identified an anomaly with Vitros system software version 3.0 & below on the Vitros 3600 immunodiagnostic system, Vitros 4600 chemistry system, and Vitros 5600 integrated system. Internal testing confirmed that under very specific conditions, incorrect patient demographics were associated with a sample identification (ID) number. This issue may occur when condition code PY1-075 is generated for a sample ID and the operator does not follow the instructions provided below in the condition review in V-docs.