Various Inline Valve Catheter and its Accessories (2014-07-17)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots
Manufacturer
Medos International SARL
Chemin-Blanc 38
Le Locle
2400
SWITZERLAND
H) Inline Valve, Siphonguard, Unit. Bact. Cath.82-8807
All lots
Manufacturer
Medos International SARL
Chemin-Blanc 38
Le Locle
2400
SWITZERLAND
F) Inline Valve, Siphonguard, Cath.,Acc.82-8805
All lots
Manufacturer
Medos International SARL
Chemin-Blanc 38
Le Locle
2400
SWITZERLAND
G) Inline Valve, Siphonguard, Unit. Cath, Acc.82-8806
All lots
Manufacturer
Medos International SARL
Chemin-Blanc 38
Le Locle
2400
SWITZERLAND
D) Inline Valve, Unit. Bactiseal Cath, Acc.82-8803
CNLCW4
Manufacturer
Medos International SARL
Chemin-Blanc 38
Le Locle
2400
SWITZERLAND
E) Inline Valve with Siphonguard82-8804
All lots
Manufacturer
Medos International SARL
Chemin-Blanc 38
Le Locle
2400
SWITZERLAND
B) Inline Valve, Catheter, Accessories82-8801
All lots
Manufacturer
Medos International SARL
Chemin-Blanc 38
Le Locle
2400
SWITZERLAND
C) Inline Valve, Unitized Catheter, Access.82-8802
More than 10 numbers, contact manufacturer.
Manufacturer
Medos International SARL
Chemin-Blanc 38
Le Locle
2400
SWITZERLAND
A) Inline Valve Only82-8800

Introduction

Recalled Products

A) Inline Valve Only
B) Inline Valve, Catheter, Accessories
C) Inline Valve, Unitized Catheter, Access.
D) Inline Valve, Unit. Bactiseal Cath, Acc.
E) Inline Valve with Siphonguard
F) Inline Valve, Siphonguard, Cath.,Acc.
G) Inline Valve, Siphonguard, Unit. Cath, Acc.
H) Inline Valve, Siphonguard, Unit. Bact. Cath.

Reason

As stated in the Codman programmable valve IFU, the virtual off setting (setting 8) is intended to limit flow through the valve and has an average pressure greater than 400mm H20. Recent In-vitro testing of the virtual off setting showed that the operating pressure for an individual valve may fall below 400mm H20. This may result in a delay in treatment for shunt-related complications.