Health Products

Vitros Chemistry Products NH3 DT Slides (2014-07-16)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
105902254396Vitros Chemistry Products NH3 DT Slides
Manufacturer
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
1532589

Introduction

Recalled Products

Vitros Chemistry Products NH3 DT Slides

Reason

OCD has determined that the data/calibration mathematics from calibration data module (CDM PROM) 0168 and 0169 for Vitros chemistry products DT calibrator kit, lot 44 required to support Vitros chemistry products NH3 DT slides, generation (GEN) 59, was inadvertently omitted. Without this calibration information, customers will be unable to calibrate or use Vitros NH3 DT slides GEN 59 with Vitros chemistry products DTcalibrator kit lot 44 when CDM PROM 0168 or 0169 is installed on their Vitros DT60/DT60II chemistry system.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime