Health Products

Vitros ECI Immunodiagnostic Products CA 19-9 Reagent Pack - Calibrator (2014-07-17)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
1320
Manufacturer
Ortho-Clinical Diagnostics
Johnson & Johnson
500-100 Holmers Farm Way
High Wycombe, Buckinghamshire
HP12 4DP
UNITED KINGDOM
Vitros ECI Immunodiagnostic Products CA 19-9 Reagent Pack – Calibrator6800035

Introduction

Recalled Products

Vitros ECI Immunodiagnostic Products CA 19-9 Reagent Pack – Calibrator

Reason

OCD internal testing has confirmed the potential for positively biased results generated when using Vitros ca 19-9 reagent pack and calibrator, lot 1320. The average bias observed is 30% for results with concentration 37 u/ml.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime