An Abbott investigation has identified that ICT Serum Calibrator lots 21399UN13, 67873UN13, 74710UNn13, 82008UN13 and 88884UN13 may generate lower than expected potassium quality control (QC) and patient results. A potential cause of this issue is increased ammonia content over the shelf life of the calibrator. Elevated ammonia concentration may impact the potassium calibration and potentially cause a calibration curve to generate lower results. Your QC and patient results recovery could vary due to the individual laboratory differences. Patient results may have shifted down by 1-3% beyond the acceptable performance of the assay.