Health Products

Steris 4085 General Surgical Table (2014-07-15)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
0418009043 to 0417114082
Manufacturer
Steris Corporation
2720 Gunter Park East
Montgomery
36109
Alabama
UNITED STATES
Steris 4085 General Surgical TableST01XXXX

Introduction

Recalled Products.

Steris 4085 General Surgical Table

Reason

Steris engineering analysis has determined the D1 pressure switch in the hydraulic column manifold does not adequately detect pressure level. If one or more of the table floor lock feet become unlocked due to the loss of hydraulic pressure, there is a potential for the table hand control to incorrectly indicate that the floor lock feet are locked.