Health Products

Buretrol Solution Set with Clearlink and Interlink (2014-07-04)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
DR13I16028
Manufacturer
Baxter Healthcare Corporation
1 Baxter Parkway
Deerfield
60015
Illinois
UNITED STATES
B. Buretrol Solution Set with Interlink (3)2C7564
DR13E03016
Manufacturer
Baxter Healthcare Corp.
One Baxter Parkway
Deerfield
60015-4625
Illinois
UNITED STATES
C. Buretrol Solution Set with Interlink (2)JC7564
DR13E08023
Manufacturer
Baxter Healthcare Corporation
1 Baxter Parkway
Deerfield
60015
Illinois
UNITED STATES
A. Buretrol Solution Set with Clearlink2C8864

Introduction

Recalled products
  1. Buretrol Solution Set with Clearlink
  2. Buretrol Solution Set with Interlink (3)
  3. Buretrol Solution Set with Interlink (2)

Reason

Baxter corporation is issuing a recall for specific lots of Buretrol solution set product codes due to complaints for separation between the burette chamber and the drip chamber. The root cause for this issue has been identified and resolved.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime