Health Products

RE Turn 7400 and 7500 (2014-04-21)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Handicare AB
Veddestavagen 15
Jarfalla
17562
SWEDEN
B. RE Turn 7500Not applicable
More than 10 numbers, contact manufacturer.
Manufacturer
Handicare AB
Veddestavagen 15
Jarfalla
17562
SWEDEN
A. RE Turn 7400Not applicable

Introduction

Recalled products
  1. RE Turn 7400
  2. RE Turn 7500

Reason

During the period April 2009 to May 2011, Handicare used a wing handle in which the screw and the handle, after repeated tightening, may slide apart. This causes a small amount of play in the attachment of the ladder, which in turn results in high load at the attachment holes of ladder. Eventually this may cause mechanical fatigue in the material of the ladder.