5mL Reaction Vessels used with the m2000sp Instrument (2014-07-09)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
56944001
Manufacturer
Abbott Molecular Inc.
1300 E. Touhy Ave.
Des Plaines
60018
Illinois
UNITED STATES
A. 5mL Reaction Vessels used with the m2000sp Instrument04J71-20

Introduction

Recalled products
  1. 5mL Reaction Vessels used with the m2000sp Instrument

Reason

During run setup on a M2000SP instrument a customer found two 5 ml reaction vessels, list number 04J71-20, with missing rims that fell through the 1ml subsystem carrier. Some of the 5 ml reaction vessels (RV) in the affected lots may have an insufficient rim.  A  RV with an insufficient rim may not be held securely in the 1ml subsystem carrier.  This may cause the RV to be seated incorrectly in the carrier or drop through it.