Siemens Healthcare Diagnostics has confirmed a carry-over risk on Sysmex CS systems due to the following applications using CA-clean I (964-0631-3) for sample probe rinse:
Protein C with protein C reagent (OQYG11)
Protein S with protein S AC reagent (OPAP03)
Factor V Leiden with proc global reagent (OQLS13)
Under certain circumstances the carry-over might impact results for Factor V with Innovin, Factor XII with Dade Actin FSL, Berichrom F XIII, Innovance D-dimer assays.
In case of Innovance D-Dimer, only frozen samples are concerned by a potential carry-over. No impact by carry-over was observed with Innovance D-Dimer when using fresh normal plasma or control plasma.