Health Products

Persona Distal Valgus Alignment Guide (2014-07-31)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
62580270
Manufacturer
Zimmer Inc.
1800 West Center Street
Warsaw
46580
Indiana
UNITED STATES
A. Persona Distal Valgus Alignment Guide42-5099-004-00

Introduction

Recalled products
  1. Persona Distal Valgus Alignment Guide

Reason

During the last 2 years, Zimmer has received 49 complaints related to binding of the Persona Distal Valgus Alignment Guide's locking knob. The root cause of the binding was determined to be excessive counter-clockwise rotation of the knob when unlocking the device. Current users are being provided with additional guidance on the locking/unlocking feature of the device.