Multiple customer complaints were received reporting HBA1C bias on patient samples and for cap survey failures when using A1C_3 calibrator lots 3HD044, 3LD068. Siemens has confirmed a -9% to -11% bias on these calibrator lots. The quality control values although high biased, recover within assigned ranges - therefore may not detect the issue. An internal complaint was also generated for NGSP certification failure due to a low bias.