Health Products

Advia 1200 Chemistry System - A1C_3 Calibrator and Advia Chemistry System - A1C_3 Calibrator (2014-08-07)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
3HD044
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
B. Advia Chemistry System – A1C_3 Calibrator10491408
3HD044
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
A. Advia 1200 Chemistry System – A1C_3 Calibrator10491408

Introduction

Recalled Products   
  1. Advia 1200 Chemistry System – A1C_3 Calibrator
  2. Advia Chemistry System – A1C_3 Calibrator

Reason

Multiple customer complaints were received reporting HBA1C bias on patient samples and for cap survey failures when using A1C_3 calibrator lots 3HD044, 3LD068. Siemens has confirmed a -9% to -11% bias on these calibrator lots. The quality control values although high biased, recover within assigned ranges - therefore may not detect the issue. An internal complaint was also generated for NGSP certification failure due to a low bias.