Columbia Agar with 5% Sheep Blood and  Trypticase Soy Agar with 5% Sheep Blood (TSA II) / MacConkey II Agar (2014-07-10)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
4143653
Manufacturer
Becton Dickinson and Company
7 Loveton Circle
Sparks
21152
Maryland
UNITED STATES
B. Trypticase Soy Agar with 5% Sheep Blood (TSA II) / MacConkey II Agar221291
4099228
Manufacturer
Becton Dickinson and Company
7 Loveton Circle
Sparks
21152
Maryland
UNITED STATES
A. Columbia Agar with 5% Sheep Blood 221263

Introduction

Recalled Products
  1. Columbia Agar with 5% Sheep Blood (2014-07-10)
  2. Trypticase Soy Agar with 5% Sheep Blood (TSA II) / MacConkey II Agar

Reason

BD has recently confirmed a low level of surface and subsurface contamination of Listeria monocytogenes on the prepared plated media. May be hazardous to health. The contamination could be erroneously reported as an actual patient result causing misdiagnosis and/or mistreatment of a non-serious illness.