Health Products

Intellispace ECG (2014-08-06)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
US31400041Intellispace ECG 
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
Massachusetts
UNITED STATES
860426

Introduction

Recalled Products

Intellispace ECG 

Reason

Philips has become aware that if multiple Quinton stress ECG records are being processed during IECG import at the same time, it is possible that the XML patient ID from one patient can be incorrectly linked to a report from a second patient that is in the processing queue.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime