Health Products

Dimension Vista System - Chem Calibrators (2014-08-08)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
3GM081Dimension Vista System - Chem Calibrators 
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
KC110

Introduction

Recalled Products   Produits retirés de marché

Dimension Vista System - Chem Calibrators 

Reason

Siemens Healthcare Diagnostics has received complaints for low recovery for magnesium (Mg) quality control (QC) and patient samples following calibration of Mg with Dimension Vista Chem 1 Cal lot 3GM081. Investigation by Siemens confirms that QC and patient samples show low recovery of Mg by 0.25 mg/dl [0.10 mmol/l] across the assay range following calibration with 3GM081.