Dimension Vista System - MMB Calibrator (2014-08-18)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
3JD045Dimension Vista System - MMB Calibrator 
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
KC672

Introduction

Recalled Products

Dimension Vista System - MMB Calibrator 

Reason

Siemens healthcare diagnostics has determined that Dimension Vista Mass CKMB Isoenzyme Calibrator (MMB Cal) shows drift during routine stability monitoring that exceeds the acceptance criteria for drift over the shelf life of this product (12 months). Siemens has observed drift from 3% to 12% at MMB concentrations across the assay range. Depending on quality control limits, this drift may not have been detected. Due to the drift, Siemens is revising the expiration date of this product (to 6 months). Siemens is issuing an urgent field safety notice letter for the aforementioned lots of Dimension Vista MMB Cal.