Health Products

Dimension Vista System-Hemoglobin A1C (HBA1C) A (2014-08-19)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
13273BA
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
A. Dimension Vista System-Hemoglobin A1C (HBA1C) A10470481

Introduction

Recalled products   
  1. Dimension Vista System-Hemoglobin A1C (HBA1C) A

Reason

An internal study has shown that variance is introduced into the percentage of HBA1C result when the sample is run from the small sample container (SSC). Previously (VC-14-03) cyclosporine was identified as under-recovering when run from the SSC. CAPA investigation has found that an unintentional dilution of the sample occurs when whole blood samples are transferred from the SSC to the aliquot well. The percentage of HBA1C is a ration of two components. With the percentage of HBA1C, the unintentional dilution of sample introduces variance into the final percentage of HBA1C result.