A. Dimension Vista System-Hemoglobin A1C (HBA1C) A
10470481
Introduction
Recalled products
Dimension Vista System-Hemoglobin A1C (HBA1C) A
Reason
An internal study has shown that variance is introduced into the percentage of HBA1C result when the sample is run from the small sample container (SSC). Previously (VC-14-03) cyclosporine was identified as under-recovering when run from the SSC. CAPA investigation has found that an unintentional dilution of the sample occurs when whole blood samples are transferred from the SSC to the aliquot well. The percentage of HBA1C is a ration of two components. With the percentage of HBA1C, the unintentional dilution of sample introduces variance into the final percentage of HBA1C result.