Health Products

STA - Liquid Anti Xa (2014-07-28)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
112366
Manufacturer
Diagnostica Stago SAS
9, Rue Des Freres Chausson
Asnieres
92600
FRANCE
STA - Liquid Anti Xa00311US

Introduction

Recalled Products

STA - Liquid Anti Xa

Reason

The Stago hotline/call center has received some customer complaints regarding a misreading of reagent 1 (substrate) barcode on STA analyzers. Diagnostica Stago has investigated and confirmed that the barcode is incorrectly coded. Consequently, when the barcode is scanned on STA analyzers, the last 3 digits displayed are "066" instead of "366". When the corresponding flyer is scanned, the STA analyzer displays an error message because the lot of the vial does not match with the lot on the flyer.