A) Fresenius 2008K Dialysate System
B) Fresenius 2008K2 Dialysate System
C) Fresenius 2008K Home
D) Fresenius 2008T Hemodialysis System
Reason
Interruption in central water delivery system resulted in a loss of water pressure causing negative pressure in water system compared to hemodialysis machine. This allowed disinfectant from machines to contaminate the central water delivery system and storage tank.