Health Products

Mobilediagnost WDR (2014-08-20)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Philips Medical Systems DMC GmbH
Rontgenstrabe 24
Hamburg
22335
GERMANY
Mobilediagnost WDR9890 010 8952X

Introduction

Mobilediagnost WDR

Reason

Philips has become aware that if the system drives over insulated flooring or carpeting and then over an uninsulated, electrically grounded section, an electrostatic discharge can cause the system's digital motion control (DMC) board to malfunction, resulting in unintended movement of the system, e.g., an increase in speed or a change of direction.