Abbott m2000 system - m2000sp and m24sp Instrument (2014-08-13)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
8055
Manufacturer
Abbott Molecular Inc.
1300 E. Touhy Ave.
Des Plaines
60018
Illinois
UNITED STATES
B) Abbott m24sp03N06-01
10040
Manufacturer
Abbott Molecular Inc.
1300 E. Touhy Ave.
Des Plaines
60018
Illinois
UNITED STATES
A) Abbott m2000 system - m2000sp Instrument9K14-02

Introduction

Recalled Products

A) Abbott m2000 system - m2000sp Instrument
B) Abbott m24sp

Reason

The X-drive assembly is used in the m2000sp and m24sp instruments.  According to X-drive assembly supplier, on one of the two processing machines where the steel wire is pretreated and primed, the wire was polluted by rusty pulleys. Consequently, belts produced from this machine between 01/01/2013 and 09/30/2013, contain rusty wires.  Under stress such as routine use of the instruments, this issue can lead to damage of the belt.  After days, weeks or sometimes months of operation the belt tension can become weak causing an unusual rattling noise or misalignment.