Health Products

Persona Knee System - Female Hex Screw (2014-09-08)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
62348078
Manufacturer
Zimmer Inc.
1800 West Center Street
Warsaw
46580
Indiana
UNITED STATES
A. Persona Knee System - Female Hex Screw42-5099-025-48

Introduction

Recalled products
  1. Persona Knee System - Female Hex Screw 

Reason

The affected product is being recalled due to potential use of the screw in cortical bone, and corresponding stripping of the 2.5mm Persona Hex Driver (42-5099-025-00).