Health Products

Bekman Coulter AU Immunochemistry System-AU480 (2013-09-03)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
2013072465
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
A. Bekman Coulter AU Immunochemistry System-AU480AU480

Introduction

Recalled products
  1. Bekman Coulter AU Immunochemistry System-AU480

Reason

When the AU480 Clinical Chemistry Analyzer operates using the real-time online function, and the number of barcode digits in the sample ID barcode label does not match the instrument configuration setting in the "requisition format" menu and in the LIS system, the AU480  Clinical Chemistry Analyzer generates alarm 6042 "online mismatch" and the sample is not processed. If the number of barcode digits configured in the LIS, middle ware, and AU480 settings match, the customer will not experience the issue. The issue does not generate incorrect results but may cause a delay in processing the sample.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime