Health Products

Aeroneb Pro AC/DC Adapter and Aeroneb Pro X Controller (2014-09-02)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable
Manufacturer
Aerogen Ltd,
Galway Business Park,
Dangan, Galway.
IRELAND
B. Aeroneb Pro X ControllerAG-PX1000-US
Not applicable.
Manufacturer
Aerogen Ltd,
Galway Business Park,
Dangan, Galway.
IRELAND
A. Aeroneb Pro AC/DC AdapterAG-AP1040-US

Introduction

Recalled products
  1. Aeroneb Pro AC/DC Adapter
  2. Aeroneb Pro X Controller

Reason

Aerogen has received a small number of complaints in relation to the mains power adapter manufactured by FRIWO GmbH and used in the Aeroneb Solo and Pro Nebuliser Systems. The complaints received identified that the power adapter cover had broken and dislodged from the body of the adapter prior to use. However it cannot be ruled out that the adapter may fail during use. If the failure were to occur during use then the internal electrical components of the mains power adapter would become exposed to the user. Therefore, there is a potential risk of electric shock if the cover falls off during use, with mains power turned on, and the user makes contact with specific recessed components inside the adapter housing.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime