Health Products

Artis Zeego Multi-Axis (2014-09-22)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
S/N 160109
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
A. Artis Zeego Multi-Axis10280959

Introduction

Recalled products
  1. Artis Zeego Multi-Axis 

Reason

  1. A cable inside the C-arm of systems affected may break within the specific serial number range of 1000 - 1132. A cable break may cause restricted functionality or even failing of the system. An ongoing procedure may be terminated at this system.
  2. A cable inside the C-arm of systems within the serial range of 1000 - 1599, may have been routed in an inappropriate manner. Without additional measure, there is potential that a cable inside the C-arm may break. The problem will not occur systematically but sporadically in the ongoing life time of a system.