Health Products

Babytherm 8000 LDR Warming Bed and Babytherm 8004/8010 (2014-08-31)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Draeger Medical GMBH
B. Babytherm 8004/8010FR 00098
More than 10 numbers, contact manufacturer.
Manufacturer
Draeger Medical GMBH
A. Babytherm 8000 LDR Warming BedFR00050

Introduction

Recalled products
  1. Babytherm 8000 LDR Warming Bed
  2. Babytherm 8004/8010

Reason

As part of Draeger's market and product monitoring they have become aware of a case in which a side panel on a Babytherm 8010 did not fully slide to the bottom in the corner guides and was not completely locked. The side panel did not open during this incident and no patient was affected. When reviewing the locking mechanism in a number of hospitals it was possible to fully close the side panels by pushing down on them slightly. However, it cannot be guaranteed that the panel will lock into position under its own weight.